Easy Step By Step Instructions
Step 1:
- Cut the PE tubing (562-2015-01) to the desired length
- Use Clean Cut Tool square off both ends of the PE tubing
Insert the tubing into the Tubing Holder (smallest hole)
Use the Distance Gauge end of Flanging Roller to set the length protruding (opposite end of flanging roller)
Lock tubing in the clamping device
Step 2:
Insert the disc (201541) with 0.80 mm pin into flanging roller
Insert disc pin into tubing end to be flanged
Apply a steady gentle pressure and rotate the flange-roller to form the flange (about 20 seconds)
Step 3:
Remove the tube from the clamping device
Install Ferrule (blue) onto the flanged tube (large end toward the flange)
Install the Nut for 4 way valve (either Red or Black as appropriate)
Step 4:
Sleeve the PE Tube ( ½ " piece 116-0536-11 Tubing) (Stretch the sleeving tube and dip into DI water)
If attaching to Glass – Cut through both tubes to form a square end then us a ¾" piece of 116-0536-16 tubing)
If attaching to Cadmium Column – the N6 nipple will connect directly into the ½ " length of 116-0536-11 tubing
Screw nut with tube attached into the 4 way valve body
|